Metal Hardness – Equostat

The Equostat features excellent sensitivity through small penetration of a few µm and is therefore particularly suited for the testing of metal hardness on scratch-sensitive and polished parts or on thin parts, profiles and pipes with a wall thickness that is below 2 mm. The Equostat was developed for accurate, rapid and reproducible metal hardness from only one side.

The Equostat is a complement to Proceq’s portable rebound metal hardness testers as it resolves most limitations associated with the dynamic testing method.

Features

  • Accurate, rapid and reproducible metal hardness results from measurement of penetration depth
  • Measurements are independent of the material and of the body dimensions
  • Testing range from very soft (aluminum) up to very hard materials (case-hardened)
  • Integrated conversions (HV, HB, HRC, HRB, HR15N, Rm)
  • Transfer of up to 5’000 stored readings to a PC via RS232 using Equolink 2
  • Excellent sensitivity down to small penetrations of only a few μm
  • The probe can be used in either a clamp, bench press or freehand
  • In addition to the R5 probe, all the impact devices of the metal hardness tester Equotip 2 and Equotip 3 can be utilized to conduct rebound hardness testing with the same indicating device

Applications

  • Quality metal hardness testing of very soft materials through to very hard materials
  • Thin parts, profiles and pipes of a wall thickness ≤ 2 mm, sheets, coiled metal, and pieces as thin as 0.2 mm (0.008”) depending on the hardness level
  • Scratch-sensitive and polished surfaces
  • Metal hardness of heat-affected zones near welding seams

Proceq offers accessories such as a hand activated clamp , a free-hand flat foot for metal hardness testing on large surfaces, support adapters for round parts, and a test stand.

Warranties

  • Standard 2-year limited warranty on electronic parts
  • Optional extended warranty periods up to 3 additional years

Standardization

  • DIN 50157
  • DGZfP Guideline “Mobile Härteprüfung“
  • VDI/VDE Guideline 2616 Paper 1

Technical Data

Equostat unit incl. R5 probe
Testing methodstatic low-load Rockwell
Available probesR5 (also Leeb rebound probes can be used)
Test loadpreload 10N, total load 50N
Conical test tipdiamond, point angle 80°
Min. radius of surface curvature1.5 mm (0.06 in.)
Penetration depth

soft materials approx. 80 μm (0.003 in.)

hard materials approx. 15 μm (0.0006 in.)
Resolution0.1 HRZ; 1 HV; 1 HB; 0.1 HRB; 0.1 HRC; 0.1 HR15N; 1 MPa (1 N/mm2)
Accuracy± 1.5 HRZ (± 2.0 HRC) throughout the entire measuring range
Equotip 2 indicating device
Dimensions175 x 180 x 80 mm (6.9 x 7.1 x 3.15 in.)
Weight750 g (incl. batteries)
Materialshock resistant ABS plastic
Displaylarge LCD
Internal data storage~ 5’000 values
CommunicationRS232 C
PC applicationEquolink 2 software included
Battery typestandard size "AA" cells (LR6) for about 65 h operation
Operating temperature0 - 50 °C (32 - 122 °F)